タイトル: | A procedure for stable electrical measurements on a rock sample against high contact resistance as a prerequisite for electrical tomography |
著者: | Suzuki, Takeshi ![]() ![]() Yoshimura, Ryokei ![]() ![]() ![]() Yamazaki, Ken’ichi Oshiman, Naoto |
著者名の別形: | 鈴木, 健士 吉村, 令慧 山﨑, 健一 大志万, 直人 |
キーワード: | Multi-point measurement Small electrode Current injection Voltage measurement Resistivity measurement Contact resistance Absolute humidity Differential method Granite Measurement environment |
発行日: | 2021 |
出版者: | Springer Nature |
誌名: | Earth, Planets and Space |
巻: | 73 |
論文番号: | 123 |
抄録: | As a basis for the electrical tomography of laboratory-scale rock samples (~ 10 cm), we developed a procedure for stable, multi-point, electrical measurement on rock samples that is effective even at high contact and sample resistance. Electrodes were strongly attached to the surface of high-resistivity rock using conductive and adhesive epoxy. Sustained current injection for long periods into high-resistance rocks was fulfilled using a constant direct current source with high internal resistance. Accurate voltage measurement across the high-resistance rock was accomplished by differential measurement using two high input resistance voltmeters. Measurements of high resistance also require a stable measurement environment: the temperature and humidity in the laboratory were controlled using an air conditioner, a humidifier, a dehumidifier, and a vinyl tent. Signal noise arising from human activities was eliminated by the remote operation of the measuring equipment and switching terminal. The proposed measurement procedure was evaluated in terms of the stability and accuracy of measured values and its applicability to electrical tomography. To assess measurement stability, we performed multiple measurements of a dry granite sample at various levels of absolute humidity. Our procedure recorded highly reproducible measurements under each humidity condition. The observed changes in measured values with absolute humidity indicate the importance of stabilising the temperature and humidity conditions in the laboratory. Applying our technique to multiple plastic samples with known resistivity confirmed its accuracy. To evaluate its applicability to electrical tomography, we measured the potential distribution on a dry granite surface in response to an injected current using a simple 40-electrode array. The potential distribution measured by our procedure agreed well with that predicted by forward modelling, demonstrating the robustness of our procedure in array measurements, and thus indicating its potential applicability to tomographic measurements for a variety of targets even under severe conditions including the relative dryness of ambient humidity. |
著作権等: | © The Author(s) 2021. This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. |
URI: | http://hdl.handle.net/2433/274608 |
DOI(出版社版): | 10.1186/s40623-021-01446-9 |
出現コレクション: | 学術雑誌掲載論文等 |

このアイテムは次のライセンスが設定されています: クリエイティブ・コモンズ・ライセンス