タイトル: 蓚酸酸化酵素を用いた尿中蓚酸測定法の検討
その他のタイトル: Evaluation of the enzymatic method using oxalate oxidase for urinary oxalate assay
著者: 小川, 由英  KAKEN_name
山口, 千美  KAKEN_name
田中, 徹  KAKEN_name
諸角, 誠人  KAKEN_name
北川, 龍一  KAKEN_name
著者名の別形: OGAWA, Yoshihide
YAMAGUCHI, Kazumi
TANAKA, Tohru
MOROZUMI, Makoto
KITAGAWA, Ryuichi
キーワード: Oxalate assay
Oxalate oxidase
Human urine
Ion chromatography
発行日: Dec-1987
出版者: 泌尿器科紀要刊行会
誌名: 泌尿器科紀要
巻: 33
号: 12
開始ページ: 1951
終了ページ: 1954
抄録: The principle of this kit method is that urinary oxalate is extracted and subsequently assayed by measuring the amount of hydrogen peroxide produced in an oxidation reaction catalyzed by oxalate oxidase. The reproducibility and accuracy of the method were tested: the within-run and day-to-day coefficients of variation were 5.4-20.0% and 16.1-18.0%, respectively, and the overall recovery rate of the added oxalate (5-25 mg/l) was 40-50%. These abnormally low recovery rates may be related to the presence of sulfate and phosphate in the extracted fluid. Therefore, the above method was modified by performing a recovery test by adding 25 mg/l oxalate to all urine samples. By the modified method, the correlation coefficient obtained between this method and the ion-chromatographic method was 0.851 (p less than 0.01). Urinary pretreatment with either acid ferric chloride or EDTA yields a higher recovery than with HCl. However, a good correlation of oxalate values is consistently observed for HCl-processed urine as measured by the above two methods. If the interference of ascorbic acid is negligible, no special urinary treatment except for HCl is necessary. The 24-hour urinary oxalate excretions were 24.4 +/- 9.1 mg (mean +/- SD) in 8 healthy males and 19.9 +/- 10.3 mg in 24 calcium-stone formers (21 males and 3 females).
URI: http://hdl.handle.net/2433/119401
PubMed ID: 3448918
出現コレクション:Vol.33 No.12

アイテムの詳細レコードを表示する

Export to RefWorks


出力フォーマット 


このリポジトリに保管されているアイテムはすべて著作権により保護されています。