タイトル: 尿中蓚酸測定キット(Sigma)の改良使用法
その他のタイトル: A modification of the estimation of urinary oxalate using a Sigma kit
著者: 戎野, 庄一  KAKEN_name
大川, 順正  KAKEN_name
著者名の別形: EBISUNO, Shoichi
OHKAWA, Tadashi
キーワード: Urinary oxalate
Oxalate oxidase
Estimation
Sigma kit
発行日: Feb-1988
出版者: 泌尿器科紀要刊行会
誌名: 泌尿器科紀要
巻: 34
号: 2
開始ページ: 225
終了ページ: 230
抄録: The Sigma kit for estimating urinary oxalate is an enzymatic procedure. However, some errors were encountered using the standard assay system of the kit. Firstly, an overestimate of oxalate may arise from the oxidation of ascorbate during the alkaline wash stage of the extraction of oxalate from urine. Secondly, an underestimate of oxalate may occur because of incomplete extraction of oxalate. A modified assay system for measurement of urinary oxalate using the kit is reported. The following points were modified: urine was diluted two-fold with 0.2 M EDTA and and 0.2 M citrate buffer (pH 3.0), oxalate from urine was extracted with 0.06 N sodium hydroxide to prevent the overestimation by the oxidation of ascorbate, and a plate mixer and addition of a small magnet to the vial were used in the steps of both absorption and extraction of oxalate to keep the accuracy of the estimation. The linearity of standard curve, reproducibility and recovery rates of the modified method were studied, and good results were obtained (linearity; r = 0.999, CV of reproducibility; 5.3%, recovery rate; 101% (300 microM) and 103% (600 microM). A good correlation was seen between the modified Sigma method and high performance liquid chromatography (r = 0.991).
URI: http://hdl.handle.net/2433/119478
PubMed ID: 3376811
出現コレクション:Vol.34 No.2

アイテムの詳細レコードを表示する

Export to RefWorks


出力フォーマット 


このリポジトリに保管されているアイテムはすべて著作権により保護されています。