ダウンロード数: 3874

このアイテムのファイル:
ファイル 記述 サイズフォーマット 
27_1019.pdf340.98 kBAdobe PDF見る/開く
タイトル: 尿流量測定の臨床的意義について
その他のタイトル: CLINICAL EVALUATION OF THE UROFLOWMETRY
著者: 八竹, 直  KAKEN_name
著者名の別形: Yachiku, Sunao
発行日: Aug-1981
出版者: 京都大学医学部泌尿器科学教室
誌名: 泌尿器科紀要
巻: 27
号: 8
開始ページ: 1019
終了ページ: 1024
抄録: One of the most important objective tests for disturbances of mictiurition is uroflowmetry (UFM), in which an apparatus (uroflowmeter) is an important factor for the determination of urinary flow curves. Six apparatus were studied about their accuracies, costs and difficulties of the operation. The mictiometer of DISA had a satisfactory accuracy and mass flowmeters (AMS & Life-Tech) could be used clinically. Flow rate nomograms for maximum flow rate and average flow rate were made as an aid in the interpretation of urinary flow data. It was shown that these nomograms were practically very useful. Urinary flow curves were classified in 4 groups. This classification was of benefit to make a diagnosis of urinary flow abnormalities. It was also demonstrated that obstructions at the bladder neck consist of two groups based on the response of urinary flow rate to phentolamine. Therefore, it would be concluded that UFM gives many informations for urodynamics in addition to a screening test of micturition.
URI: http://hdl.handle.net/2433/122936
出現コレクション:Vol.27 No.8

アイテムの詳細レコードを表示する

Export to RefWorks


出力フォーマット 


このリポジトリに保管されているアイテムはすべて著作権により保護されています。