ダウンロード数: 362

このアイテムのファイル:
ファイル 記述 サイズフォーマット 
s00586-011-1728-3.pdf3.39 MBAdobe PDF見る/開く
タイトル: A novel synthetic material for spinal fusion: a prospective clinical trial of porous bioactive titanium metal for lumbar interbody fusion
著者: Fujibayashi, Shunsuke  KAKEN_id
Takemoto, Mitsuru
Neo, Masashi
Matsushita, Tomiharu
Kokubo, Tadashi
Doi, Kenji
Ito, Tatsuya  KAKEN_id
Shimizu, Akira  kyouindb  KAKEN_id
Nakamura, Takashi
著者名の別形: 藤林, 俊介
キーワード: Porous titanium metal
Spinal fusion
Biomaterial
Clinical trial
発行日: Sep-2011
出版者: Springer-Verlag
誌名: European spine journal
巻: 20
号: 9
開始ページ: 1486
終了ページ: 1495
抄録: The objective of this study was to establish the efficacy and safety of porous bioactive titanium metal for use in a spinal fusion device, based on a prospective human clinical trial. A high-strength spinal interbody fusion device was manufactured from porous titanium metal. A bioactive surface was produced by simple chemical and thermal treatment. Five patients with unstable lumbar spine disease were treated surgically using this device in a clinical trial approved by our Ethics Review Committee and the University Hospital Medical Information Network. Clinical and radiological results were reported at the minimum follow-up period of 1 year. The optimal mechanical strength and interconnected structure of the porous titanium metal were adjusted for the device. The whole surface of porous titanium metal was treated uniformly and its bioactive ability was confirmed before clinical use. Successful bony union was achieved in all cases within 6 months without the need for autologous iliac crest bone grafting. Two specific findings including an anchoring effect and gap filling were evident radiologically. All clinical parameters improved significantly after the operation and no adverse effects were encountered during the follow-up period. Although a larger and longer-term follow-up clinical study is mandatory to reach any firm conclusions, the study results show that this porous bioactive titanium metal is promising material for a spinal fusion device.
著作権等: The final publication is available at www.springerlink.com
This is not the published version. Please cite only the published version.
この論文は出版社版でありません。引用の際には出版社版をご確認ご利用ください。
URI: http://hdl.handle.net/2433/152450
DOI(出版社版): 10.1007/s00586-011-1728-3
PubMed ID: 21369760
出現コレクション:学術雑誌掲載論文等

アイテムの詳細レコードを表示する

Export to RefWorks


出力フォーマット 


このリポジトリに保管されているアイテムはすべて著作権により保護されています。