タイトル: Dosimetric advantages of O-ring design radiotherapy system for skull-base tumors
著者: Ogura, Kengo
Mizowaki, Takashi  kyouindb  KAKEN_id  orcid https://orcid.org/0000-0002-8135-8746 (unconfirmed)
Ishida, Yuichi
Hiraoka, Masahiro  KAKEN_id
著者名の別形: 溝脇, 尚志
キーワード: dynamic conformal arc radiotherapy
dosimetric comparison
O-ring design radiotherapy system
skull-base tumors
fractionated stereotactic radiotherapy
発行日: 2014
出版者: American Association of Physicists in Medicine
誌名: Journal of Applied Clinical Medical Physics
巻: 15
号: 2
開始ページ: 226
終了ページ: 234
抄録: The purpose of this study was to investigate whether a new O-ring design radiotherapy delivery system has advantages in radiotherapy planning for skull-base tumors. Twenty-five patients with skull-base tumors were included in this study. Two plans were made using conventional (Plan A) or new (Plan B) techniques. Plan A consisted of four dynamic conformal arcs (DCAs): two were horizontal, and the other two were from cranial directions. Plan B was created by converting horizontal arcs to those from caudal directions making use of the O-ring design radiotherapy system. The micromultileaf collimators were fitted to cover at least 99% of the planning target volume with prescribed doses, 90% of the dose at the isocenter. The two plans were compared in terms of target homogeneity, conformity, and irradiated volume of normal tissues, using a two-sided paired t-test. For evaluation regarding target coverage, the homogeneity indices defined by the International Commission on Radiation Units and Measurements 83 were 0.099± 0.010 (mean± standard deviation) and 0.092 ± 0.010, the conformity indices defined by the Radiation Therapy Oncology Group were 1.720 ± 0.249 and 1.675± 0.239, and the Paddick’s conformity indices were 0.585 ± 0.078 and 0.602 ± 0.080, in Plans A and B, respectively. For evaluation of irradiated normal tissue, the Paddick’s gradient indices were 3.118 ± 0.283 and 2.938 ± 0.263 in Plans A and B, respectively. All of these differences were statistically significant (p-values < 0.05). The mean doses of optic nerves, eyes, brainstem, and hippocampi were also significantly lower in Plan B. The DCA technique from caudal directions using the new O-ring design radiotherapy system can improve target homogeneity and conformity compared with conventional DCA techniques, and can also decrease the volume of surrounding normal tissues that receives moderate doses.
著作権等: This work is licensed under a Creative Commons Attribution 3.0 License.
URI: http://hdl.handle.net/2433/187098
DOI(出版社版): 10.1120/jacmp.v15i2.4608
PubMed ID: 24710448
関連リンク: http://www.jacmp.org/index.php/jacmp/article/view/4608/pdf_59
出現コレクション:学術雑誌掲載論文等

アイテムの詳細レコードを表示する

Export to RefWorks


出力フォーマット 


このリポジトリに保管されているアイテムはすべて著作権により保護されています。