Access count of this item: 220

Files in This Item:
File Description SizeFormat 
j.aca.2014.11.016.pdf328.19 kBAdobe PDFView/Open
Title: An off-line automated preconcentration system with ethylenediaminetriacetate chelating resin for the determination of trace metals in seawater by high-resolution inductively coupled plasma mass spectrometry.
Authors: Minami, Tomoharu
Konagaya, Wataru
Zheng, Linjie
Takano, Shotaro  kyouindb  KAKEN_id
Sasaki, Masanobu
Murata, Rena
Nakaguchi, Yuzuru
Sohrin, Yoshiki  kyouindb  KAKEN_id
Author's alias: 宗林, 由樹
Keywords: Trace metals
Seawater
Automated preconcentration
Chelating resin
Inductively coupled plasma mass spectrometry
Issue Date: 7-Jan-2015
Publisher: Elsevier B.V.
Journal title: Analytica chimica acta
Volume: 854
Start page: 183
End page: 190
Abstract: A novel automated off-line preconcentration system for trace metals (Al, Mn, Fe, Co, Ni, Cu, Zn, Cd, and Pb) in seawater was developed by improving a commercially available solid-phase extraction system SPE-100 (Hiranuma Sangyo). The utilized chelating resin was NOBIAS Chelate-PA1 (Hitachi High-Technologies) with ethylenediaminetriacetic acid and iminodiacetic acid functional groups. Parts of the 8-way valve made of alumina and zirconia in the original SPE-100 system were replaced with parts made of polychlorotrifluoroethylene in order to reduce contamination of trace metals. The eluent pass was altered for the back flush elution of trace metals. We optimized the cleaning procedures for the chelating resin column and flow lines of the preconcentration system, and developed a preconcentration procedure, which required less labor and led to a superior performance compared to manual preconcentration (Sohrin et al.). The nine trace metals were simultaneously and quantitatively preconcentrated from ∼120 g of seawater, eluted with ∼15 g of 1M HNO3, and determined by HR-ICP-MS using the calibration curve method. The single-step preconcentration removed more than 99.998% of Na, K, Mg, Ca, and Sr from seawater. The procedural blanks and detection limits were lower than the lowest concentrations in seawater for Mn, Ni, Cu, and Pb, while they were as low as the lowest concentrations in seawater for Al, Fe, Co, Zn, and Cd. The accuracy and precision of this method were confirmed by the analysis of reference seawater samples (CASS-5, NASS-5, GEOTRACES GS, and GD) and seawater samples for vertical distribution in the western North Pacific Ocean.
Rights: © 2014 Elsevier B.V.
This is not the published version. Please cite only the published version. この論文は出版社版でありません。引用の際には出版社版をご確認ご利用ください。
URI: http://hdl.handle.net/2433/193261
DOI(Published Version): 10.1016/j.aca.2014.11.016
PubMed ID: 25479883
Appears in Collections:Journal Articles

Show full item record

Export to RefWorks


Export Format: 


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.