dc.contributor.authorTanimura, Kazuyaen
dc.contributor.authorHirai, Toyohiroen
dc.contributor.authorSato, Susumuen
dc.contributor.authorHasegawa, Koichien
dc.contributor.authorMuro, Shigeoen
dc.contributor.authorKurosawa, Hajimeen
dc.contributor.authorMishima, Michiakien
dc.contributor.alternative平井, 豊博ja
dc.date.accessioned2015-10-08T05:20:34Z-
dc.date.available2015-10-08T05:20:34Z-
dc.date.issued2014-07-18-
dc.identifier.issn1880-6562-
dc.identifier.urihttp://hdl.handle.net/2433/200214-
dc.description.abstractSince commercial forced oscillation technique (FOT) devices became available, they have been widely used for physiological assessments, mainly of obstructive lung diseases. However, it is not known whether the impedance values measured with different devices are identical. In this study, two FOT devices-the impulse oscillometry system (IOS) and the MostGraph (MG)-were compared using phantom models. The resistance values varied up to 10% from estimated values in both devices. Additionally, there was a difference in frequency dependence for the resistance between the devices. The reactance values measured with MG were higher than those measured with IOS. The effects of ventilation on the measured impedance values were higher for IOS than for MG, especially at lower frequencies. We concluded that the devices do not always generate identical impedance values. Thus, differences between the devices should be taken into consideration when evaluating clinical data.en
dc.format.mimetypeapplication/pdf-
dc.language.isoeng-
dc.publisherSpringer Japanen
dc.rightsThe final publication is available at Springer via http://dx.doi.org/10.1007/s12576-014-0329-4.en
dc.rightsこの論文は出版社版でありません。引用の際には出版社版をご確認ご利用ください。ja
dc.rightsThis is not the published version. Please cite only the published version.en
dc.subjectForced oscillationen
dc.subjectImpedanceen
dc.subjectResistanceen
dc.subjectReactanceen
dc.subjectFrequency dependenceen
dc.subject.meshAirway Resistance/physiologyen
dc.subject.meshElectric Impedanceen
dc.subject.meshModels, Biologicalen
dc.subject.meshOscillometry/methodsen
dc.subject.meshPulmonary Ventilation/physiologyen
dc.subject.meshRespirationen
dc.titleComparison of two devices for respiratory impedance measurement using a forced oscillation technique: basic study using phantom models.en
dc.typejournal article-
dc.type.niitypeJournal Article-
dc.identifier.jtitleThe journal of physiological sciencesen
dc.identifier.volume64-
dc.identifier.issue5-
dc.identifier.spage377-
dc.identifier.epage382-
dc.relation.doi10.1007/s12576-014-0329-4-