dc.contributor.author | Asao, Yasufumi | en |
dc.contributor.author | Hashizume, Yohei | en |
dc.contributor.author | Suita, Takahiro | en |
dc.contributor.author | Nagae, Ken Ichi | en |
dc.contributor.author | Fukutani, Kazuhiko | en |
dc.contributor.author | Sudo, Yoshiaki | en |
dc.contributor.author | Matsushita, Toshikazu | en |
dc.contributor.author | Kobayashi, Shuichi | en |
dc.contributor.author | Tokiwa, Mariko | en |
dc.contributor.author | Yamaga, Iku | en |
dc.contributor.author | Fakhrejahani, Elham | en |
dc.contributor.author | Torii, Masae | en |
dc.contributor.author | Kawashima, Masahiro | en |
dc.contributor.author | Takada, Masahiro | en |
dc.contributor.author | Kanao, Shotaro | en |
dc.contributor.author | Kataoka, Masako | en |
dc.contributor.author | Shiina, Tsuyoshi | en |
dc.contributor.author | Toi, Masakazu | en |
dc.contributor.alternative | 髙田, 正泰 | ja |
dc.contributor.alternative | 片岡, 正子 | ja |
dc.contributor.alternative | 椎名, 毅 | ja |
dc.contributor.alternative | 戸井, 雅和 | ja |
dc.date.accessioned | 2017-02-24T06:25:18Z | - |
dc.date.available | 2017-02-24T06:25:18Z | - |
dc.date.issued | 2016-11-23 | - |
dc.identifier.issn | 1560-2281 | - |
dc.identifier.uri | http://hdl.handle.net/2433/218411 | - |
dc.description.abstract | We have constructed a prototype photoacoustic mammography system (PAM-02) capable of simultaneously acquiring photoacoustic (PA) and ultrasound (US) images. Each PA, US, and fused PA/US image can be acquired over a wide area of the breast using the scanning module of a US transducer, a PA detector, and optical prisms. The resolution of the PA images exhibits improvement from 2 to 1 mm compared to images acquired using our previous prototype. The maximum scan area of PAM-02 is 90 mm along the horizontal axis and 150 mm along the vertical axis. In a phantom experiment, the available depth was at least 45 mm. A representative example of the application of the PAM-02 prototype in clinical research at Kyoto University is presented and shows S-factor images, which are considered an approximation parameter related to hemoglobin saturation of tumor-related blood vessels. We confirmed the applicability of the system for anatomical and biological research. | en |
dc.format.mimetype | application/pdf | - |
dc.language.iso | eng | - |
dc.publisher | SPIE | en |
dc.rights | © The Authors. Published by SPIE under a Creative Commons Attribution 3.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI. | en |
dc.subject | photoacoustic imaging | en |
dc.subject | optoacoustic | en |
dc.subject | mammography | en |
dc.subject | ultrasound | en |
dc.title | Photoacoustic mammography capable of simultaneously acquiring photoacoustic and ultrasound images | en |
dc.type | journal article | - |
dc.type.niitype | Journal Article | - |
dc.identifier.jtitle | Journal of Biomedical Optics | en |
dc.identifier.volume | 21 | - |
dc.identifier.issue | 11 | - |
dc.relation.doi | 10.1117/1.JBO.21.11.116009 | - |
dc.textversion | publisher | - |
dc.identifier.artnum | 116009 | - |
dc.identifier.pmid | 27893089 | - |
dcterms.accessRights | open access | - |
出現コレクション: | 学術雑誌掲載論文等 |

このリポジトリに保管されているアイテムはすべて著作権により保護されています。