dc.description.abstractThe purpose of this study was to develop a simple verification method for the routine quality assurance (QA) of Dynamic WaveArc (DWA) irradiation using electronic portal imaging device (EPID) images and log data analysis. First, an automatic calibration method utilizing the outermost multileaf collimator (MLC) slits was developed to correct the misalignment between the center of the EPID and the beam axis. Moreover, to verify the detection accuracy of the MLC position according to the EPID images, various positions of the MLC with intentional errors in the range 0.1–1 mm were assessed. Second, to validate the geometric accuracy during DWA irradiation, tests were designed in consideration of three indices. Test 1 evaluated the accuracy of the MLC position. Test 2 assessed dose output consistency with variable dose rate (160–400 MU/min), gantry speed (2.2–6°/s), and ring speed (0.5–2.7°/s). Test 3 validated dose output consistency with variable values of the above parameters plus MLC speed (1.6–4.2 cm/s). All tests were delivered to the EPID and compared with those obtained using a stationary radiation beam with a 0° gantry angle. Irradiation log data were recorded simultaneously. The 0.1‐mm intentional error on the MLC position could be detected by the EPID, which is smaller than the EPID pixel size. In Test 1, the MLC slit widths agreed within 0.20 mm of their exposed values. The averaged root‐mean‐square error (RMSE) of the dose outputs was less than 0.8% in Test 2 and Test 3. Using log data analysis in Test 3, the RMSE between the planned and recorded data was 0.1 mm, 0.12°, and 0.07° for the MLC position, gantry angle, and ring angle, respectively. The proposed method is useful for routine QA of the accuracy of DWA.en
dc.format.mimetypeapplication/pdf-
dc.language.isoeng-
dc.publisherWileyen
dc.rights© 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.en
dc.subjectDynamic WaveArcen
dc.subjectelectronic portal imaging deviceen
dc.subjectlog dataen
dc.subjectquality assuranceen
dc.titleQuality assurance of geometric accuracy based on an electronic portal imaging device and log data analysis for Dynamic WaveArc irradiationen
dc.typejournal article-
dc.type.niitypeJournal Article-
dc.identifier.jtitleJournal of Applied Clinical Medical Physicsen
dc.identifier.volume19-
dc.identifier.issue3-
dc.identifier.spage234-
dc.identifier.epage242-
dc.relation.doi10.1002/acm2.12324-
dc.textversionpublisher-
dc.addressDepartment of Radiation Oncology and Image‐applied therapy, Graduate School of Medicine, Kyoto Universityen
dc.addressDepartment of Radiation Oncology and Image‐applied therapy, Graduate School of Medicine, Kyoto Universityen
dc.addressDepartment of Radiation Oncology and Image‐applied therapy, Graduate School of Medicine, Kyoto Universityen
dc.addressDepartment of Radiation Oncology and Image‐applied therapy, Graduate School of Medicine, Kyoto Universityen
dc.addressDepartment of Radiation Oncology and Image‐applied therapy, Graduate School of Medicine, Kyoto Universityen
dc.addressDepartment of Radiation Oncology and Image‐applied therapy, Graduate School of Medicine, Kyoto Universityen
dc.identifier.pmid29633542-
dcterms.accessRightsopen access-
出現コレクション:学術雑誌掲載論文等

アイテムの簡略レコードを表示する

Export to RefWorks


出力フォーマット 


このリポジトリに保管されているアイテムはすべて著作権により保護されています。