dc.description.abstractThe response of a 1-inch diameter by 6-inch long ³He proportional counter, of which the gas filling is 4 atm ³He and 2 atm krypton, to axially incident neutrons has been studied and expressed as a 50 by 50 matrix for an energy range from zero to 5 MeV. Applications of the matrix were examined for the purpose of obtaining response-corrected spectra of neutrons transmitted through a matter.en
dc.language.isoeng-
dc.publisherFaculty of Engineering, Kyoto Universityen
dc.publisher.alternative京都大学工学部ja
dc.subject.ndc500-
dc.titleResponse Function of a 1-inch Diameter by 6-inch Long ³He Proportional Counter to Neutronsen
dc.typedepartmental bulletin paper-
dc.type.niitypeDepartmental Bulletin Paper-
dc.identifier.ncidAA00732503-
dc.identifier.jtitleMemoirs of the Faculty of Engineering, Kyoto Universityen
dc.identifier.volume35-
dc.identifier.issue3-
dc.identifier.spage309-
dc.identifier.epage329-
dc.textversionpublisher-
dc.sortkey08-
dc.addressDepartment of Nuclear Engineeringen
dc.addressDepartment of Nuclear Engineeringen
dc.addressDepartment of Nuclear Engineeringen
dcterms.accessRightsopen access-
dc.identifier.pissn0023-6063-
出現コレクション:Vol.35 Part 3

アイテムの簡略レコードを表示する

Export to RefWorks


出力フォーマット 


このリポジトリに保管されているアイテムはすべて著作権により保護されています。