タイトル: Automatic measurement of beam‐positioning accuracy at off‐isocenter positions
著者: Ono, Tomohiro
Kido, Takahisa
Nakamura, Mitsuhiro
Iramina, Hiraku
Kakino, Ryo
Mizowaki, Takashi
著者名の別形: 小野, 智博
木戸, 貴久
中村, 光宏
伊良皆, 拓
溝脇, 尚志
キーワード: beam-positioning accuracy
developer mode
off isocenter
Winston–Lutz test
発行日: Mar-2023
出版者: Wiley
The American Association of Physicists in Medicine
誌名: Journal of Applied Clinical Medical Physics
巻: 24
号: 3
論文番号: e13844
抄録: [Purpose] This study performed an automatic measurement of the off-axis beam-positioning accuracy at a single isocenter via the TrueBeam Developer mode and evaluated the beam-positioning accuracy considering the effect of couch rotational errors. [Methods] TrueBeam STx and the Winston–Lutz test-dedicated phantom, with a 3 mm diameter steel ball, were used in this study. The phantom was placed on the treatment couch, and the Winston–Lutz test was performed at the isocenter for four gantry angles (0°, 90°, 180°, and 270°) using an electronic portal imaging device. The phantom offset positions were at distances of 0, 25, 50, 75, and 100 mm from the isocenter along the superior–inferior, anterior–posterior, and left–right directions. Seventeen patterns of multileaf collimator-shaped square fields of 10 × 10 mm² were created at the isocenter and off-axis positions for each gantry angle. The beam-positioning accuracy was evaluated with couch rotation along the yaw-axis (0°, ± 0.5°, and ± 1.0°). [Results] The mean beam-positioning errors at the isocenter and off-isocenter distances (from the isocenter to ±100 mm) were 0.46–0.60, 0.44–0.91, and 0.42–1.11 mm for the couch angles of 0°, ±0.5°, and ±1°, respectively. The beam-positioning errors increased as the distance from the isocenter and couch rotation increased. [Conclusion] These findings suggest that the beam-positioning accuracy at the isocenter and off-isocenter positions can be evaluated quickly and automatically using the TrueBeam Developer mode. The proposed procedure is expected to contribute to an efficient evaluation of the beam-positioning accuracy at off-isocenter positions.
著作権等: © 2022 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine.
This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
URI: http://hdl.handle.net/2433/287022
DOI(出版社版): 10.1002/acm2.13844
PubMed ID: 36420973
出現コレクション:学術雑誌掲載論文等

アイテムの詳細レコードを表示する

Export to RefWorks


出力フォーマット 


このアイテムは次のライセンスが設定されています: クリエイティブ・コモンズ・ライセンス Creative Commons