このアイテムのアクセス数: 4

このアイテムのファイル:
ファイル 記述 サイズフォーマット 
jrr_rrad086.pdf1.53 MBAdobe PDF見る/開く
完全メタデータレコード
DCフィールド言語
dc.contributor.authorHirashima, Hideakien
dc.contributor.authorNakamura, Mitsuhiroen
dc.contributor.authorNakamura, Kiyonaoen
dc.contributor.authorMatsuo, Yukinorien
dc.contributor.authorMizowaki, Takashien
dc.date.accessioned2025-06-13T01:00:08Z-
dc.date.available2025-06-13T01:00:08Z-
dc.date.issued2024-01-
dc.identifier.urihttp://hdl.handle.net/2433/294647-
dc.description.abstractThe applications of Type B [anisotropic analytical algorithm (AAA) and collapsed cone (CC)] and Type C [Acuros XB (AXB) and photon Monte Carlo (PMC)] dose calculation algorithms in spine stereotactic body radiotherapy (SBRT) were evaluated. Water- and bone-equivalent phantoms were combined to evaluate the percentage depth dose and dose profile. Subsequently, 48 consecutive patients with clinical spine SBRT plans were evaluated. All treatment plans were created using AXB in Eclipse. The prescription dose was 24 Gy in two fractions at a 10 MV FFF on TrueBeam. The doses were then recalculated with AAA, CC and PMC while maintaining the AXB-calculated monitor units and beam arrangement. The dose index values obtained using the four dose calculation algorithms were then compared. The AXB and PMC dose distributions agreed with the bone-equivalent phantom measurements (within ±2.0%); the AAA and CC values were higher than those in the bone-equivalent phantom region. For the spine SBRT plans, PMC, AAA and CC were overestimated compared with AXB in terms of the near minimum and maximum doses of the target and organ at risk, respectively; the mean dose difference was within 4.2%, which is equivalent with within 1 Gy. The phantom study showed that the results from AXB and PMC agreed with the measurements within ±2.0%. However, the mean dose difference ranged from 0.5 to 1 Gy in the spine SBRT planning study when the dose calculation algorithms changed. Users should incorporate a clinical introduction that includes an awareness of these differences.en
dc.language.isoeng-
dc.publisherOxford University Press (OUP)en
dc.publisherThe Japanese Radiation Research Societyen
dc.publisherJapanese Society for Radiation Oncologyen
dc.rights© The Author(s) 2023. Published by Oxford University Press on behalf of The Japanese Radiation Research Society and Japanese Society for Radiation Oncology.en
dc.rightsThis is an Open Access article distributed under the terms of the Creative Commons Attribution License, whichpermits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited.en
dc.rights.urihttps://creativecommons.org/licenses/by/4.0/-
dc.subjectdosimetric verificationen
dc.subjectdose calculation algorithmen
dc.subjectspine SBRTen
dc.titleDosimetric verification of four dose calculation algorithms for spine stereotactic body radiotherapyen
dc.typejournal article-
dc.type.niitypeJournal Article-
dc.identifier.jtitleJournal of Radiation Researchen
dc.identifier.volume65-
dc.identifier.issue1-
dc.identifier.spage109-
dc.identifier.epage118-
dc.relation.doi10.1093/jrr/rrad086-
dc.textversionpublisher-
dc.identifier.pmid37996097-
dcterms.accessRightsopen access-
dc.identifier.eissn1349-9157-
出現コレクション:学術雑誌掲載論文等

アイテムの簡略レコードを表示する

Export to RefWorks


出力フォーマット 


このアイテムは次のライセンスが設定されています: クリエイティブ・コモンズ・ライセンス Creative Commons