ダウンロード数: 333

このアイテムのファイル:
ファイル 記述 サイズフォーマット 
v6-p04.pdf324.58 kBAdobe PDF見る/開く
タイトル: Development of Gas Scintillation Proportional Counter (STATES AND STRUCTURES-Atomic and Molecular Physics)
著者: Shigeoka, Nobuyuki
Mutaguchi, Kohei
Nakanishi, Yoshikazu
Ito, Yoshiaki
Mukoyama, Takeshi
キーワード: Gas Scintillation
Proportional counter
Mn K X-ray
Pulse-height distribution
発行日: Mar-2000
出版者: Institute for Chemical Research, Kyoto University
誌名: ICR Annual Report
巻: 6
開始ページ: 4
終了ページ: 5
抄録: Properties of gas scintillation proportional counter are investigated for Mn K X-ray spectra. The energy resolution (8.9%) on this type of the counter has twice as good as that (19%) of Xe gas sealed proportional counter. In the counter, a distortion of the pulse-height distribution due to a low-energy tail is observed. This fact can be attributed to the penetration of high energy X-rays in the accelerating region. Therefore, it is considerable that the tail may be enhanced for higher energy photons.
URI: http://hdl.handle.net/2433/65248
出現コレクション:Vol.6 (1999)

アイテムの詳細レコードを表示する

Export to RefWorks


出力フォーマット 


このリポジトリに保管されているアイテムはすべて著作権により保護されています。