タイトル: | Development of Gas Scintillation Proportional Counter (STATES AND STRUCTURES-Atomic and Molecular Physics) |
著者: | Shigeoka, Nobuyuki Mutaguchi, Kohei Nakanishi, Yoshikazu Ito, Yoshiaki Mukoyama, Takeshi |
キーワード: | Gas Scintillation Proportional counter Mn K X-ray Pulse-height distribution |
発行日: | Mar-2000 |
出版者: | Institute for Chemical Research, Kyoto University |
誌名: | ICR Annual Report |
巻: | 6 |
開始ページ: | 4 |
終了ページ: | 5 |
抄録: | Properties of gas scintillation proportional counter are investigated for Mn K X-ray spectra. The energy resolution (8.9%) on this type of the counter has twice as good as that (19%) of Xe gas sealed proportional counter. In the counter, a distortion of the pulse-height distribution due to a low-energy tail is observed. This fact can be attributed to the penetration of high energy X-rays in the accelerating region. Therefore, it is considerable that the tail may be enhanced for higher energy photons. |