dc.description.abstractCharacteristics of the capacitive hygrometer have been investigated. The errors of relativehumidity (RH) are less than + - 3% RH and the time constant is 1 sec at a temperature of 20C.A long-term drift has been found, but is small and less than 3% RH per year. However, thelinearity of the output of hygrometer has not been changed.From a comparative observation between the capacitive hygrometer and the infrared hygro-meter, it is found that the frequency response function of the capacitive hygrometer to RHfluctuations is different from that expected from a first order device.en
dc.format.mimetypeapplication/pdf-
dc.language.isojpn-
dc.publisher京都大学防災研究所ja
dc.publisher.alternativeDisaster Prevention Research Institute, Kyoto Universityen
dc.subject.ndc451.35-
dc.title容量型湿度計の特性についてja
dc.title.alternativeCHARACTERISTICS OF THE CAPACITIVE HYGROMETERen
dc.typedepartmental bulletin paper-
dc.type.niitypeDepartmental Bulletin Paper-
dc.identifier.ncidAN00027784-
dc.identifier.jtitle京都大学防災研究所年報. Bja
dc.identifier.volume31-
dc.identifier.issueB-1-
dc.identifier.spage183-
dc.identifier.epage191-
dc.textversionpublisher-
dc.sortkey16-
dc.relation.urlhttp://www.dpri.kyoto-u.ac.jp/nenpo/nenpo.html-
dcterms.accessRightsopen access-
dc.identifier.pissn0386-412X-
dc.identifier.jtitle-alternativeDisaster Prevention Research Institute Annuals. Ben
出現コレクション:No.31 B-1

アイテムの簡略レコードを表示する

Export to RefWorks


出力フォーマット 


このリポジトリに保管されているアイテムはすべて著作権により保護されています。