タイトル: 新しい尿失禁防止具(Penoring仮称)の試み
その他のタイトル: CLINICAL APPLICATION OF A NEW TYPE OF A PENILE CLAMP (PENORING) TO PREVENT MALE URINARY INCONTINENCE
著者: 平賀, 聖悟  KAKEN_name
内島, 豊  KAKEN_name
岡田, 耕市  KAKEN_name
駒瀬, 元治  KAKEN_name
上田, 忠和  KAKEN_name
横川, 正之  KAKEN_name
水尾, 敏之  KAKEN_name
牛山, 武久  KAKEN_name
武田, 裕寿  KAKEN_name
著者名の別形: Hiraga, Seigo
Uchijima, Yutaka
Okada, Koichi
Komase, Motoharu
Ueda, Tadakazu
Yokokawa, Masayuki
Mizuo, Toshiyuki
Ushiyama, Takehisa
Takeda, Hirohisa
発行日: Aug-1981
出版者: 京都大学医学部泌尿器科学教室
誌名: 泌尿器科紀要
巻: 27
号: 8
開始ページ: 967
終了ページ: 974
抄録: A new type of penile clamp (Penoring) was applied to 14 male urinary incontinence patients including 5 post-prostatectomy incontinence and 9 neurogenic bladder dysfunction. The device is made from dimethylpolysiloxane, which is ring-shaped and is placed along the penile shaft from the glans penis. A small balloon is attached to the inner surface of the ring. The balloon can be inflated or deflated through a silicone tube connected to a small pinch-pump. It selectively compresses the urethra of the patient when inflated to prevent urinary incontinence. The clamp was evaluated not only with respect to changes of complaints but also through x-ray examinations and/or urethral pressure profilometry (UPP) before and after application of the Penoring. Urinary incontinence was successfully prevented in II of 14 patients. Maximum urethral pressure averaged 63cmH2O without the Penoring and increased to more than 129cm H2O with the device according to UPP measurement. Functional profile length of average 5.4 cm increased to 11.8 cm after the application of the device. Urethrogram or voiding cysto-urethrogram aslo showed its effectiveness in preventing urinary incontinence. The Penoring was found to have various advantages. It is light, non-irritating to the foreskin and makes possible to change the clamp effect by regulating the pressure of the ba lloon according to the degree of the incontinence. It proved to be applicable even in small boys. However, it was obliged to discontinue its use in 3 patients and several other patients were not satisfied enough with the device so they used the Penoring in conjunction with previous methods for controlling urinary incontinence. The main cause of these problems was thought to be that the size of silicone ring in the Penoring did not exactly match the penile shaft of each patients so as to get a complete clamp effect on the urethra. Some redesigning of the device, therefore, is intended in the future.
URI: http://hdl.handle.net/2433/122943
出現コレクション:Vol.27 No.8

アイテムの詳細レコードを表示する

Export to RefWorks


出力フォーマット 


このリポジトリに保管されているアイテムはすべて著作権により保護されています。