タイトル: | Improving measurement accuracy by optimum data acquisition for Nd:YAG Thomson scattering systema) |
著者: | Minami, T. Itoh, Y. Yamada, I. Yasuhara, R. Funaba, H. Nakanishi, H. Hatae, T. |
著者名の別形: | 南, 貴司 |
発行日: | Nov-2014 |
出版者: | AIP Publishing |
誌名: | Review of Scientific Instruments |
巻: | 85 |
号: | 11 |
論文番号: | 11D837 |
抄録: | A new high speed Nd:YAG Thomson scattering AD Convertor (HYADC) that can directly convert the detected scattered light signal into a digital signal is under development. The HYADC is expected to improve a signal to noise ratio of the Nd:YAG Thomson scattering measurement. The data storage of the HYADC which is required for the direct conversion of whole plasma discharge is drastically reduced by a ring buffer memory and a stop trigger system. Data transfer of the HYADC is performed by the SiTCP. The HYADC is easily expandable to a multi-channel system by the distributed data processing, and is very compact and easy to implement as a built-in system of the polychromators. |
著作権等: | © 2014 AIP Publishing LLC |
URI: | http://hdl.handle.net/2433/189551 |
DOI(出版社版): | 10.1063/1.4891416 |
PubMed ID: | 25430250 |
出現コレクション: | 学術雑誌掲載論文等 |