dc.contributor.authorMukumoto, Nobutakaen
dc.contributor.authorNakamura, Mitsuhiroen
dc.contributor.authorSawada, Akiraen
dc.contributor.authorSuzuki, Yasunobuen
dc.contributor.authorTakahashi, Kunioen
dc.contributor.authorMiyabe, Yukien
dc.contributor.authorKaneko, Shujien
dc.contributor.authorMizowaki, Takashien
dc.contributor.authorKokubo, Masakien
dc.contributor.authorHiraoka, Masahiroen
dc.contributor.alternative中村, 光宏ja
dc.date.accessioned2014-09-24T05:52:16Z-
dc.date.available2014-09-24T05:52:16Z-
dc.date.issued2013-04-
dc.identifier.issn0094-2405-
dc.identifier.urihttp://hdl.handle.net/2433/189832-
dc.description.abstract[Purpose] To verify the accuracy of an infrared (IR) marker-based dynamic tumor-tracking irradiation system (IR tracking) using the gimbaled x-ray head of the Vero4DRT (MHI-TM2000). Conclusions: Positional tracking errors correlated strongly with 4D-modeling errors in IR tracking. Thus, the accuracy of the 4D model must be verified before treatment, and margins are required to compensate for the 4D-modeling error [Conclusions] Positional tracking errors correlated strongly with 4D-modeling errors in IR tracking. Thus, the accuracy of the 4D model must be verified before treatment, and margins are required to compensate for the 4D-modeling error.en
dc.format.mimetypeapplication/pdf-
dc.language.isoeng-
dc.publisherAmerican Association of Physicists in Medicineen
dc.rights© 2013 American Association of Physicists in Medicineen
dc.rightsThis is not the published version. Please cite only the published version.en
dc.rightsこの論文は出版社版でありません。引用の際には出版社版をご確認ご利用ください。ja
dc.subject.meshEquipment Designen
dc.subject.meshEquipment Failure Analysisen
dc.subject.meshFiducial Markersen
dc.subject.meshHumansen
dc.subject.meshImaging, Three-Dimensional/instrumentationen
dc.subject.meshInfrared Rays/diagnostic useen
dc.subject.meshNeoplasms/diagnosisen
dc.subject.meshNeoplasms/radiotherapyen
dc.subject.meshPattern Recognition, Automated/methodsen
dc.subject.meshPhantoms, Imagingen
dc.subject.meshRadiotherapy, Image-Guided/instrumentationen
dc.subject.meshReproducibility of Resultsen
dc.subject.meshSensitivity and Specificityen
dc.subject.meshThermography/instrumentationen
dc.subject.meshTomography, X-Ray Computed/instrumentationen
dc.titleAccuracy verification of infrared marker-based dynamic tumor-tracking irradiation using the gimbaled x-ray head of the Vero4DRT (MHI-TM2000).en
dc.typejournal article-
dc.type.niitypeJournal Article-
dc.identifier.ncidAA00727671-
dc.identifier.jtitleMedical physicsen
dc.identifier.volume40-
dc.identifier.issue4-
dc.relation.doi10.1118/1.4794506-
dc.textversionauthor-
dc.identifier.artnum41706-
dc.identifier.pmid23556876-