タイトル: | Accuracy verification of infrared marker-based dynamic tumor-tracking irradiation using the gimbaled x-ray head of the Vero4DRT (MHI-TM2000). |
著者: | Mukumoto, Nobutaka https://orcid.org/0000-0002-4569-3379 (unconfirmed) Nakamura, Mitsuhiro Sawada, Akira Suzuki, Yasunobu Takahashi, Kunio Miyabe, Yuki Kaneko, Shuji Mizowaki, Takashi https://orcid.org/0000-0002-8135-8746 (unconfirmed) Kokubo, Masaki Hiraoka, Masahiro |
著者名の別形: | 中村, 光宏 |
発行日: | Apr-2013 |
出版者: | American Association of Physicists in Medicine |
誌名: | Medical physics |
巻: | 40 |
号: | 4 |
論文番号: | 41706 |
抄録: | [Purpose] To verify the accuracy of an infrared (IR) marker-based dynamic tumor-tracking irradiation system (IR tracking) using the gimbaled x-ray head of the Vero4DRT (MHI-TM2000). Conclusions: Positional tracking errors correlated strongly with 4D-modeling errors in IR tracking. Thus, the accuracy of the 4D model must be verified before treatment, and margins are required to compensate for the 4D-modeling error [Conclusions] Positional tracking errors correlated strongly with 4D-modeling errors in IR tracking. Thus, the accuracy of the 4D model must be verified before treatment, and margins are required to compensate for the 4D-modeling error. |
著作権等: | © 2013 American Association of Physicists in Medicine This is not the published version. Please cite only the published version. この論文は出版社版でありません。引用の際には出版社版をご確認ご利用ください。 |
URI: | http://hdl.handle.net/2433/189832 |
DOI(出版社版): | 10.1118/1.4794506 |
PubMed ID: | 23556876 |