タイトル: | Quantitative Susceptibility Mapping at 3 T and 1.5 T: Evaluation of Consistency and Reproducibility. |
著者: | Hinoda, Takuya ![]() ![]() Fushimi, Yasutaka ![]() ![]() ![]() Okada, Tomohisa ![]() ![]() Fujimoto, Koji ![]() ![]() ![]() Liu, Chunlei Yamamoto, Akira ![]() ![]() Okada, Tsutomu ![]() Kido, Aki Togashi, Kaori ![]() |
著者名の別形: | 日野田, 卓也 伏見, 育崇 |
キーワード: | magnetic resonance imaging |
発行日: | 18-Apr-2015 |
出版者: | Wolters Kluwer Health, Inc |
誌名: | Investigative radiology |
巻: | 50 |
号: | 8 |
開始ページ: | 522 |
終了ページ: | 530 |
抄録: | [Objectives]: The aim of this study was to assess the consistency and reproducibility of quantitative susceptibility mapping (QSM) at 3-T and 1.5-T magnetic resonance (MR) scanners. [Materials and Methods]: This study was approved by institutional ethics committee, and written informed consent was obtained. Twenty-two healthy volunteers underwent 2 examinations on different days. Each examination consisted of MR imaging on both 3-T and 1.5-T MR scanners. The data from both scanners and examination days were obtained, and QSM was calculated with STI Suite using 2 different algorithms-harmonic phase removal using laplacian operator (HARPERELLA) and a sophisticated harmonic artifact reduction for phase data (SHARP) method with a variable radius of the spherical kernel at the brain boundary (V-SHARP). We evaluated consistency of QSM between 3 T and 1.5 T and the reproducibility between the first and second examinations using 2-phase processing methods (HARPERELLA and V-SHARP). [Results]: Susceptibility values of regions of interests at 3 T were highly correlated with those at 1.5 T with good agreement (HARPERELLA, R2 = 0.838; V-SHARP, R2 = 0.898) (average difference, +/-1.96 SD; HARPERELLA, -0.012 +/- 0.046; V-SHARP, -0.002 +/- 0.034). Reproducibility analysis demonstrated excellent correlation between the first and second examination at both 3 T and 1.5 T for both algorithms (HARPERELLA at 3 T, R2 = 0.921; 1.5 T, R2 = 0.891; V-SHARP at 3 T, R2 = 0.937; 1.5 T, R2 = 0.926). Bland-Altman analysis showed excellent reproducibility for HARPERELLA (3 T, -0.003 +/- 0.032; 1.5 T, -0.003 +/- 0.038) and V-SHARP (3 T, -0.003 +/- 0.027; 1.5 T, -0.003 +/- 0.029). Susceptibility values of these 2 algorithms were highly correlated with good agreement (3T, R2 = 0.961; 1.5 T, R2 = 0.931) (3 T, 0.009 +/- 0.023; 1.5 T, -0.003 +/- 0.049). [Conclusions]: Quantitative susceptibility mapping with HARPERELLA and V-SHARP demonstrated good reproducibility at 3 T and 1.5 T, and QSM with V-SHARP demonstrated good consistency at 3 T and 1.5 T. |
著作権等: | © 2015 Wolters Kluwer Health. This is a non-final version of an article published in final form in Investigative Radiology: Post Author Corrections: April 18, 2015 doi: 10.1097/RLI.0000000000000159. 許諾条件により本文ファイルは2016-04-18に公開. この論文は出版社版でありません。引用の際には出版社版をご確認ご利用ください。 This is not the published version. Please cite only the published version. |
URI: | http://hdl.handle.net/2433/197533 |
DOI(出版社版): | 10.1097/RLI.0000000000000159 |
PubMed ID: | 25900085 |
出現コレクション: | 学術雑誌掲載論文等 |