タイトル: | Photoacoustic mammography capable of simultaneously acquiring photoacoustic and ultrasound images |
著者: | Asao, Yasufumi Hashizume, Yohei Suita, Takahiro Nagae, Ken Ichi Fukutani, Kazuhiko Sudo, Yoshiaki Matsushita, Toshikazu Kobayashi, Shuichi Tokiwa, Mariko Yamaga, Iku Fakhrejahani, Elham Torii, Masae Kawashima, Masahiro https://orcid.org/0000-0002-4738-8351 (unconfirmed) Takada, Masahiro https://orcid.org/0000-0002-5954-1296 (unconfirmed) Kanao, Shotaro Kataoka, Masako https://orcid.org/0000-0002-6212-3351 (unconfirmed) Shiina, Tsuyoshi Toi, Masakazu |
著者名の別形: | 髙田, 正泰 片岡, 正子 椎名, 毅 戸井, 雅和 |
キーワード: | photoacoustic imaging optoacoustic mammography ultrasound |
発行日: | 23-Nov-2016 |
出版者: | SPIE |
誌名: | Journal of Biomedical Optics |
巻: | 21 |
号: | 11 |
論文番号: | 116009 |
抄録: | We have constructed a prototype photoacoustic mammography system (PAM-02) capable of simultaneously acquiring photoacoustic (PA) and ultrasound (US) images. Each PA, US, and fused PA/US image can be acquired over a wide area of the breast using the scanning module of a US transducer, a PA detector, and optical prisms. The resolution of the PA images exhibits improvement from 2 to 1 mm compared to images acquired using our previous prototype. The maximum scan area of PAM-02 is 90 mm along the horizontal axis and 150 mm along the vertical axis. In a phantom experiment, the available depth was at least 45 mm. A representative example of the application of the PAM-02 prototype in clinical research at Kyoto University is presented and shows S-factor images, which are considered an approximation parameter related to hemoglobin saturation of tumor-related blood vessels. We confirmed the applicability of the system for anatomical and biological research. |
著作権等: | © The Authors. Published by SPIE under a Creative Commons Attribution 3.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI. |