タイトル: Potential utility of cone‐beam CT‐guided adaptive radiotherapy under end‐exhalation breath‐hold conditions for pancreatic cancer
著者: Ogawa, Ayaka
Nakamura, Mitsuhiro
Iramina, Hiraku
Yoshimura, Michio
Mizowaki, Takashi
著者名の別形: 小川, 彩加
中村, 光宏
伊良皆, 拓
吉村, 通央
溝脇, 尚志
キーワード: CBCT-guided adaptive radiotherapy
dosimetric analysis
independent calculation-based QA
pancreatic cancer
発行日: Feb-2023
出版者: Wiley
The American Association of Physicists in Medicine
誌名: Journal of Applied Clinical Medical Physics
巻: 24
号: 2
論文番号: e13827
抄録: [Purpose] The purpose of this study was to demonstrate the potential utility of cone-beam computed tomography (CBCT)-guided online adaptive radiotherapy (ART) under end-exhalation breath-hold (EE-BH) conditions for pancreatic cancer (PC). [Methods] Eleven PC patients who underwent 15-fraction volumetric-modulated arc therapy under EE-BH conditions were included. Planning CT images and daily 165 CBCT images were imported into a dedicated treatment planning system. The prescription dose was set to 48 Gy in 15 fractions. The reference plan was automatically generated along with predefined clinical goals. After segmentation was completed on CBCT images, two different plans were generated: One was an adapted (ADP) plan in which re-optimization was performed on the anatomy of the day, and the other was a scheduled (SCH) plan, which was the same as the reference plan. The dose distributions calculated using the synthetic CT created from both planning CT and CBCT were compared between the two plans. Independent calculation-based quality assurance was also performed for the ADP plans, with a gamma passing rate of 3%/3 mm. [Results] All clinical goals were successfully achieved during the reference plan generation. Of the 165 sessions, gross tumor volume D98% and clinical target volume D98% were higher in 100 (60.1%) and 122 (74.0%) ADP fractions. In each fraction, the V3 Gy < 1 cm³ of the stomach and duodenum was violated in 47 (28.5%) and 48 (29.1%), respectively, of the SCH fractions, whereas no violations were observed in the ADP fractions. There were statistically significant differences in the dose–volume indices between the SCH and ADP fractions (p < 0.05). The gamma passing rates were above 95% in all ADP fractions. [Conclusions] The CBCT-guided online ART under EE-BH conditions successfully reduced the dose to the stomach and duodenum while maintaining target coverage.
著作権等: © 2022 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine.